OPERATOR'S MANUAL
®
mini GAC
11001-1464 Rev C

NORMAL OPERATION

The Home menu will appear immediately after all startup self
tests successfully complete. This is the "Home" menu through
which all other menus are accessed.

MEASURING MOISTURE

In preparation for testing grain, the following conditions should
be observed.
1. The cell MUST be empty and clean prior to testing for
accurate, consistent measurements.
2. The Loader MUST be removed before turning the unit On
and during all testing (hour bar indicates when testing),
refer to
(Figure
IMPORTANT:
The loader should only be in or on the unit
during storage and when grain is loaded.
The Home menu will display the active grain at the top of the
Home screen.
SELECTING GRAIN
To select a new grain:
1. Press the Down Arrow button to position the cursor at the
Product heading and press Enter.
2. Select a product (grain) using the Up or Down Arrow
button (to check a grain's calibration constant, remain on a
selected grain for approximately 3 seconds and the
calibration constant menu will display. Pressing the Down
Arrow button will scroll to the next Constant menu and the
Bias menu) or immediately press the Enter or Home
button to save and change the selection.
After grain selection, the new product (grain) will display at the
top of the Home menu. The grain list on the grain selection menu
will be saved in a "last used" order.
IMPORTANT:
Periodic updates of the calibration constants
are recommended. Reference the Calibration
Grain Values section for additional
information.
15).
NORMAL OPERATION / 15