5. Instructions for Use
P1 Regular
FILLING
35 min
P1 Regular
Cycle Complete
P1 Regular
Instruments
not processed
Cycle Aborted
OK
HYDR IM C51wd Operator's Manual
The cycle will start by draining the unit for 20 seconds. The bar
on the screen indicates the progression of the cycle. The stage
within the cycle is shown beneath the bar (eg. Filling, Rinsing).
The estimated time remaining (in minutes) is always displayed.
Please note that the time will increase or decrease depending
on the projected cycle completion.
When the cycle is complete, the program name and "Cycle
Complete" is displayed, and the unit beeps three times. Press
to open the door. Use gloves to unload the instruments. After all
cycles, exercise caution as the instruments, trolley, baskets and
cassettes may be hot. Inspect the instruments for debris prior to
packaging / sterilising.
5.5 Interrupting / Aborting a Cycle
The door of the Hydrim is locked during the cycle. If the STOP
button is pressed, the cycle is aborted and cannot be re-started.
The message "Draining. Please wait" will appear. The unit will
drain. Then the message "Instruments not processed. Cycle
Aborted" will appear. Press OK. Open the door by pressing
Use gloves if handling the instruments. Exercise caution as
instruments, trolley, baskets and cassettes may be hot. The
cycle must be repeated to ensure that instruments are properly
processed. Note that if the drying phase of the cycle is
interrupted, the critical parts of the cycle are complete and the
message displaying the program name and "Cycle Complete.
Drying interrupted" will appear.
.
Page 10