Setting the temperature control mode

When using an external PT 1000 temperature sensor, the user can
choose between two types of control:
PI mode
Good control results, minimized overshooting, slow rise in temperature
2P-mode (two-point controller)
Maximum heating rate, increased overshooting
Factory setting: PI mode
Changing the temperature control mode
1. Set target temperature to 2 °C
2. Switch off device
3. Hold down temp knob
4. Switch on device
This activates or deactivates the two-point controller (2P) for the
external PT 1000 temperature sensor (2P appears on the display).
When switching on the device, 2P is also displayed next to the
operating mode A/B/D when the two-point controller is activated.
When the heating function is active in 2P mode, (with external PT 1000
temperature sensor) the actual temperature flashes - that notifies the
user that the process temperature has been greatly overshot!

Maintenance and cleaning

The device is maintenance-free.
Cleaning
- For cleaning disconnect the mains plug!
- Use only cleaning agents which have been approved by IKA to
clean IKA devices.
These are water (with tenside) and isopropanol.
- Wear protective gloves during cleaning the devices.
24
- Electrical devices may not be placed in the cleansing agent for
the purpose of cleaning.
- Do not allow moisture to get into the device when cleaning.
- Before using another than the recommended method for
cleaning or decontamination, the user must ascertain with IKA
that this method does not destroy the device.
Ordering spare parts
When ordering spare parts, please give:
- Machine type
- Manufacturing number, see type plate
- Item number and designation of the spare part,
see www.ika.com.
Repair
Please send instrument in for repair only after it has been
cleaned and is free from any materials which may constitu-
te a health hazard.
For this you should request the "Safety Declaration (Deconta-
mination Certificate)" from IKA, or use the download printout
of it from the IKA website www.ika.com.
Return the instrument in its original packaging. Storage packaging is
not sufficient. Also, please use suitable shipping package materials.

Accessories

• For further accessories see www.ika.com.
Tabla de contenido
loading

Este manual también es adecuado para:

0003810000

Tabla de contenido