Operation
Calibration of buffer 2 starts 2 sec after pressing the keys CAL, "UP" or
"DOWN", as above; marker "P2" is displayed steadily.
After successful calibration the LCD elements "P1" and "P2" are switched
off and the new calibrating data is displayed:
The meter display returns to the measuring operation previous to the
calibration. If the temperature was being measured, the parameter pH is
selected.
6.4.2
Press the key "CAL" followed by the key "pH/mV/°C" to display the
calibrating data again later.
The meter display returns to the measuring operation previous to the
calibration. If the temperature was being measured, the parameter pH is
selected.
6.4.3
Press the key "pH/mV/°C" after successfully calibrating buffer 1 to
discontinue the calibration process.
The calibration is now interpreted as a 1-point calibration and new
calibrating data are displayed. Depending on the value of the selected
buffer, the zero point or slope are calibrated:
Selected buffer value:
New calculation of:
36
zero point for 2 sec with activated LCD elements "cal" and "∆pH", and
afterwards
slope for 2 sec with activated LCD elements "cal" and "%".
Displaying calibration data
One point calibration
selected buffer value
new calculation of:
0.00 – 5.49
slope
5.50 – 8.50
8.51 – 14.00
zero point
slope
ProMinent
®