To clear error press the standby button
Problem
Cause
Unit fails to power on
Missing or blown fuse
Rattling or ticking sounds may
Unit is excessively noisy
indicate a loose screw on the tray
RTD open or over temperature
E1
above 100ºC
RTD shorted or temperature
E2
below 0ºC
Drive system failure
Loose suction cup foot
E3
Ceased bearing
Drive belt broken
Unit overload or
E4
Loose suction cup foot
E5
Unit temperature is above 80˚C
10
Solution
Add or replace fuse as necessary. If problem persists, please contact your Ohaus Representative for repair.
Ensure that tray is secured tightly. If problem persists, please contact your Ohaus Representative for repair.
This error cannot be fixed by end user. Please contact your Ohaus Representative for repair.
This error cannot be fixed by end user. Please contact your Ohaus Representative for repair.
Remove the mechanical obstruction. In the event a foot (suction cup) has come loose from the bench top, firmly
press down on the four (4) corners of the unit creating a strong suction to the work surface (DO NOT place on
bench mat). If problem persists, please contact your Ohaus Representative for repair.
Ensure the load is evenly distributed and does not exceed the maximum load capacity for the unit. In the event
a foot (suction cup) has come loose from the bench top, firmly press down on the four (4) corners of the unit
creating a strong suction to the work surface (DO NOT place on bench mat). If problem persists, please contact
your Ohaus Representative for repair.
Press the standby button to turn the unit off and allow it to cool. If the E5 error persists, switch the unit off and
contact your Ohaus representative for repair.