NOTE: The latest edition of this operating manual can be downloaded from our website at www.maxtec.com
WARRANTY
Under normal operating conditions, Maxtec warrants the Acrylic Flowmeters to be free from
defects of workmanship or materials for the following period:
a) Acrylic Housing ...........................................................................................Lifetime of the product
b) Needle Valve ........................................................................................Five (5) years from receipt
c) All other parts not listed ......................................................................Two (2) years from receipt
Warranty does not cover breakage/abuse.
These warranties are from the date of receipt, provided that the product is properly oper-
ated and maintained in accordance with Maxtec's operating instructions. Based on Maxtec
product evaluation, Maxtec's sole obligation under foregoing warranty is limited to making
replacements, repairs, or issuing credit for equipment found to be defective. This warranty
extends only to the buyer purchasing the equipment directly from Maxtec or through Maxtec's
designated distributors and agents as new equipment. Routine maintenance items, such as
o-rings, are excluded from warranty. Maxtec and any other subsidiaries shall not be liable
to the purchaser or other persons for incidental or consequential damages or equipment that
has been subject to abuse, misuse, mis-application, alteration, negligence or accident. These
warranties are exclusive and in lieu of all other warranties, expressed or implied, including
warranty of merchantability and fitness for a particular purpose.
For product warranty returns, please contact Maxtec Customer Service for a Returned Materials
Authorization (RMA).
WARNINGS
This manual instructs a professional to install and operate the flowmeter. This is provided for
your safety and to prevent damage to the flowmeter. If you do not understand this manual,
DO NOT USE the flowmeter and contact your provider.
Use flowmeters only for their "Intended Use" as described in this manual.
ALWAYS confirm prescribed flow before administering to patient and monitor flow on a fre-
quent basis.
Flowmeters may contain magnetic, ferrous material that may affect the results of an MRI.
Read this User Manual before installing or operating the flowmeter.
To Reduce the Risk of Fire or Explosion:
ALWAYS follow ANSI and CGA standards for Medical Gas Products and flowmeters and
Oxygen Handling.
DO NOT use or store oils, greases, organic lubricants or any combustible materials on or
near this flowmeter.
DO NOT use near any type of flame or flammable/explosive substances, vapors or
atmosphere.
DO NOT connect to source pressure greater than 100 psi.
ENGLISH
phone: (800) 748.5355
fax: (801) 973.6090
web: www.maxtec.com
AUTHORIZED REPRESENTATIVE:
QNET BV, Kantstraat 19, NL-5076 NP Haaren, The Netherlands
DO NOT disassemble the flowmeter while pressurized.
DO NOT smoke in an area where oxygen is being administered.
CAUTIONS
Use caution to avoid cross-threading fittings during installation.
Always use a backing wrench when installing or removing connectors.
Flowmeters must be operated with the Flow Tube in a vertical, upright position.
Only personnel instructed and trained in its use should operate this flowmeter.
Ensure all connections are tight and leak free.
Only use oxygen-safe leak detector.
Only trained medical professionals should operate the flowmeter.
DO NOT drop the flowmeter.
DO NOT autoclave.
DO NOT gas sterilize with EtO (Ethylene Oxide).
DO NOT clean with aromatic hydrocarbons.
DO NOT immerse flowmeter in any kind of liquid. This will void the warranty.
DO NOT use if damaged. Inspect the flowmeter for visual damage before each use.
DO NOT over-tighten knob when turning off. This will cause damage to the flowmeter.
DO NOT over-tighten any threaded connectors; body may crack and cause leaks.
The flowmeter has been specifically calibrated for use on the outlet of an air/oxygen blender.
The accuracy of the flow rates will be affected if the flowmeter is used in any other way
Gas temperatures other than 70° F (21°C) may affect the accuracy of the indicated flow.
Attaching accessories to the outlet (which may increase resistance to outlet flow) may
change the indicated flow but will not affect the accuracy of the flow.
2
WWW .MAXTEC .COM • (800) 748-5355