eNviroNMeNtal coNDitioNs
For indoor use only
Operating range
Storage range
Degree of pollution
Altitude
reFereNce coNDitioNs
Ambient temperature
Relative humidity
Atmospheric pressure
Frequency of electrical network
FUNctioNal cHaracteristics
electrical characteristics
Nominal range : 5 A
Protection against permanent overload up to 10 A
Output/Input ratio : 0.2 mV AC /mA AC
Primary voltage loss :< 0.3 V
accuracy to be added to the ca82XX, ca83XX, ca84XX and poWerpaD accuracy
Current RMS
5 mA; 50 mA
50 mA; 6 A
intrinsic accuracy tu be added to pel 1XX accuracy
Current RMS
5 mA; 250 mA
250 mA; 6 A
errors of influence
Frequency influence :
Frequency influence
Uncertainty
Phase error
Temperature influence :< 0.1 % / ° K
coNstrUctioN speciFicatioNs
Dimensions (L x W X H)
Leads
Plugs
Weight
Protection index
coMpliaNce WitH iNterNatioNal staNDarDs
Application of safety rules as per IEC 61010-1.
Pollution type 2
Installation category III and service voltage 300V
speciFicatioNs
-15 °C to +55 °C and 10 % to 90 % RH
-40 °C to +85 °C and 10 % to 90 % RH
2
< 2000 m
23 °C +/- 3K
[45 % ; 75 %]
[860 HPA ; 1060 HPa]
50 Hz +/- 0.1 Hz and 60 Hz +/- 0.1 Hz
65 Hz to 500 Hz
0.1 %
0.1 °
200 x 12 x 80 mm
- one lead 1.7 m long and 16 mm in diameter
- four shielded leads 1 metre long
- specific ESSAILEC® current input test block
- specific to the C.A 82XX, CA83XX, C.A 84XX and PEL1XX voltage output plugs
approximately 2.4 kg
IP50 for the housing and IP 20 for the terminals, according to IEC 60529
RMS
Uncertainty
± (1 %)
± (0.5 %)
Uncertainty
± 0.5 % ± 2 mA
± 0.5 % ± 1 mA
500 Hz to 1 kHz
0.3 %
0.2 °
in relation to earth, double isolation.
27
Phase error
± (1°)
± (0°)
Phase error
-
± 0.5°
1 kHz to 5 kHz
0.5 %
1 °