3.2 COMPLIANCE ENHANCEMENT SETTINGS
Somnibel
The
device includes additional features for enhancing patient compliance to the positional therapy.
These features are configured with default parameters, as follows:
Default
Settings
value
EasySleep
15 min
Delay
Supine Delay
30 sec
Stimulus
Increase
60 sec
Period
Time at
Maximum
20 min
Power
Back to Sleep
5 min
Delay
EN 13
Description
Time between device ON and activation of stimulus if
patient is in supine position.
Time between detection of supine position and activation
of stimulus.
Time between consecutive increases of stimulus
intensity.
Maximum time with the stimulus at 100% of intensity if
patient does not change position.
When the device detects the upright position and returns
to any other position, it waits some minutes before
applying again the stimulus.
somnibel
User's Manual
Enhancement
Allows the patient to fall sleep at the
beginning of treatment.
Allows the patient to transition between
left and right positions without applying
any stimulus.
Modifies the rate of stimulus increase
for adapting it to each patient's needs.
Avoids excessive application of
vibration stimulus.
Allows the patient to fall sleep again
after getting up in the middle of the
night.