Replacing Parts - OHAUS VF Serie Manual De Instrucciones

Ocultar thumbs Ver también para VF Serie:
Tabla de contenido
Idiomas disponibles
  • ES

Idiomas disponibles

VF Series Floor Scales

9. REPLACING PARTS

9.1 Replacing the Analogue PCB
1)
Set scale up on edge.
2)
Remove the analogue PCB:
Unscrew and remove the cover of the connection box.
Detach the connecting wires of the load cells and the connecting cables to the indicator.
Mark the cables for subsequent reconnection.
Detach the PCB from the retainer and remove it.
3)
Reinstall the analogue PCB:
Connect the load cells to the analogue PCB by attaching the load cell connecting cable to the
corresponding terminal strip in accordance with the following table:
Terminal
EXC +
SIG +
SIG -
EXC -
Attach the connecting cable to the indicator in accordance with the following table:
Terminal
EXC +
SEN +
SIG +
SIG -
SEN -
EXC -
Note: The load cells are calibrated at the factory in such a way that adjustment of the corner load is
usually not required. If corrections are required in individual cases, adjustment can be carried out at the
potentiometer pairs 1, 2, 3 and 4.
State
Colour
Supply +
Green
Signal +
White
Signal -
Red
Supply -
Black
State
Colour
Supply +
Grey
Control line +
Yellow
Signal +
White
Signal -
Brown
Control line -
Green
Supply -
Blue
EN-7