• Do not bend flexible instrument handles beyond a 90° angle.
• All instruments must be sterilized in a tray or container that is
permeable to steam.
• Do not expose instruments to cleaning or rinse agents that are not
compatible with polyphenylsulfone.
DIRECTIONS FOR USE
Press the straight, sterilized handle into the mechanical valve holder/
rotator receptacle. Verify that the handle is seated securely.
Screw the sterilized extension handle model EX-05 or EX2000-R into
the sterilized handle model MHV-HHR, MHV-HHL1, MHV-HHL2,
or MHV-HHL3 (Figure 1). Verify that the extension handle is securely
attached.
Figure 1
NOTE: When using the handle/extension handle combination
to rotate mechanical valves, clockwise rotation is recommended.
Counter-clockwise rotation could unscrew the extension handle
from the base handle.