CAUTION: Sterilization temperatures must not exceed
138°C/280°F.
NOTE: Higher temperatures or longer sterilization times may be
used; however; increasing the sterilization time or temperature (not to
exceed 138°C/280°F) of the stated cycles may reduce instrument life.
Temperature
Time
DISASSEMBLY AFTER USE
Pull on the handles to disengage them from the mechanical valve
holder/rotators.
Unscrew the handles from the 905-35 and 905-37 sizers, or the EX-05
and EX2000-R extension handles.
Handles must be disassembled prior to cleaning and sterilization.
DISPOSAL
The instructions for use is recyclable. Dispose of all packaging
materials as appropriate. Dispose of accessories per standard solid
biohazard waste procedures.
LIMITED WARRANTY
The user must comply with the terms and conditions of this limited
warranty in order for this limited warranty to apply.
Pre-Vac Sterilization Parameters
10
134°C/273°F
5 minutes