J. Test for true rotation of levitated test
impeller by moving the switch on the control
box to the ON position and turn the RPM
Regulator Knob clock-wise and set the arrow
at 50 (See Fig. 18). The levitated impeller
will spin. The levitation gap must remain
uniform (no wobbling of the impeller)
K. Turn the Regulator Knob to 0 and move
the switch on the control box to the OFF
position and then remove the test impeller.
Do not try to remove the test impeller while
it is spinning (See Fig. 18). Replace the pro-
tective shield on the test impeller. The unit
is now ready to be positioned under the
®
®
Palletank
for LevMixer
or Palletank
for LevMixer
®
to carry out the mixing.
Figure 18.
Cautions:
Always put the protective shield back on test
impeller when testing procedure is complete.
1
If the test impeller does not levitate or
substantial wobbling takes place (more
than 1mm change in gap) the system
must be reset. Turn the system off for
25 minutes and repeat set up procedure.
If problem continues, contact a technical
representative of Sartorius Stedim Biotech.

5.3. Preparation of the Palletank

LevMixer
®
or Palletank
LevMixer
®
A. Before proceeding to the Flexel
®
for LevMixer
installation in the Palletank
check that the internal surfaces of the
1
.
®
Palletank
are smooth, free of any particles
that might damage the bag surface and dry
(See Fig. 19).
®
jacketed
Figure 19.
B. Apply the Palletank
C. Open the hinged front door by the locking
systems and remove the front 4-port system
gate by sliding it up from its position
(See 1 and 3 of Fig. 20).
D. Position the clamp holder on top of the
Palletank
®
and slide its two parts open.
(See 4 of Fig. 20)
E. Remove the centering aligner
(See 2 of Fig. 20).
Figure 20.
®
for
5.4. Flexel
®
Jacketed for
in the Palletank
Palletank
®
Bag
A. Obtain a packaged Flexel
®
,
LevMixer
matches Palletank
smooth table with no sharp angular corners.
B. Check that the irradiation indicator on
the label is red. Record the batch number
of the bag (as seen on the product label).
(See Fig. 21)
Dry
Clean
No particle
Figure 21.
C. To open and remove the secondary
(external) packaging (See Fig. 22):
- cut one of its corner just under the closure
®
two brakes
weld with a pair of rounded scissors and
open completely by tearing along the weld.
- remove the secondary packaging by sliding
it away from the Flexel
D. Check that the Flexel
secondary packaging is free from any
damage that could compromise the sterility.
E. Repeat the C), D) steps for the primary
(internal) packaging.
®
®
Bag for LevMixer
installation
®
for LevMixer
®
or
®
Jacketed for LevMixer
®
®
Bag for
®
of the appropriate size (Bag size
®
size) and position it on a
®
Bag for LevMixer
®
Bag for LevMixer
®
.
®
11