8.4  E rror limits
Message
Out of range, determine again
Standard temp. out of range
ATC measured temperature is different
to the user-defined value
Temperature out of range
Temp. out of nLF correction range
Temp. out of pure water range
Temp. out of conductivity ash cor r ec-
tion range
40 Maintenance
Description and Resolution
The exporting process failed. This could be due to the fol-
low i ng reasons:
The USB stick is not connected or it is disconnected
dur i ng the exporting process
The USB stick is full
Range not accepted
Conductivity
TDS
Salinity
Resistivity
Conductivity ash
Temperature
| tATC-Tstandard | > 1 °C
Temperature
Temperature
Temperature
Temperature
< 0.00 µS/cm or > 1000
mS/cm
< 0.00 mg/L or > 600 g/L
< 0.00 psu or > 80.0 psu
< 0.00 MΩ•cm or > 100.0
MΩ•cm
< 0.00 % or > 2022 %
< 0 °C or > 35 °C
< -5 °C or > 105 °C
< 0°C or > 50C
< 0 °C or > 50 °C
< 15 °C or > 25 °C