Message
Active sensor cannot be deleted
Wrong standard
Standard temp. out of range
Temperature differs from setting
ISM
sensor communication error
®
Self-test failure
Wrong settings
Out of range
38 Maintenance
Description and Resolution
Deleting the calibration data of the selected sensor ID is
not possible, because it is currently the active sensor ID
in the meter shown on the display.
Enter new sensor ID in the menu settings.
Select another sensor ID from the list in the menu
set t ings.
Meter cannot recognize the standard.
Make sure that you have the correct standard and that it
is fresh.
The ATC measured temperature is out of calibration
standard range: 5 ... 35 °C for international standards
and 15 ... 35°C for Chinese standards
Keep the standard temperature within the range.
Change the temperature setting.
ATC measured temperature differs by more than 0.5°C
from the user-defined value/temperature range.
Keep the standard temperature within the range.
Change the temperature setting.
Data has not been transferred correctly between ISM
sen s or and meter. Reconnect the ISM
again.
Self-test has not been completed within 2 minutes or me-
ter is defective.
Restart self-test and finish within 2 minutes.
Contact METTLER TOLEDO service if problem per-
sists.
Entered value differs by less than 5°C from other pre-
set val u es.
Enter a higher/lower value in order to get a bigger
dif f er e nce.
Either entered value is out of range.
Enter a value which is within the range shown on
dis p lay.
or
Measured value out of range.
Make sure the electrode wetting cap has been re-
moved and that the electrode is properly connected
and placed in the sample solution.
®
sensor and try
®