Performance Specifications; Physical Specifications - AND UA-779 Manual De Instrucciones

Tabla de contenido
Idiomas disponibles

Idiomas disponibles

Performance Specifications

Measurement method Oscillometric measurement
Measurement range
Battery
Cuff
Classification
Clinical test
Environment Specifications
Operating environment
Storage environment

Physical Specifications

Dimensions
Weight
Specifications
Pressure : 20 - 280 mmHg ±4 mmHg
Pulse :
40 - 200 beats / minute ±5 %
4 x 1.5V (R6P, LR6, or AA)
Arm Circumference
Type BF 0366
Device, cuff and tubing are designed to provide special
protection against electric shocks.
According to ANSI / AAMI SP-10 1987
Systolic blood pressure
Diastolic blood pressure
+10 °C - +40 °C, less than 85 %RH
-20 °C - +70 °C, less than 85 %RH
142 [W] x 131 [D] x 67 [H] mm
Approx. 330g without batteries
ENGLISH 11
Standard deviation
5.43 mmHg
5.18 mmHg
Mean error
2.24 mmHg
1.54 mmHg