3.5.3 Connect a power cord—Analyzer without enclosure
Note: Do not use conduit to supply power.
6
Item supplied by user: Power cord
1. Remove the electrical access cover. Refer to
Remove the electrical access cover
on page 15.
2. Connect a power cord. Refer to the illustrated steps that follow.
3. Install the electrical access cover.
4. Do not connect the power cord to an electrical outlet.
6
Refer to
Power cord guidelines
on page 21.