Electrical interference immunity
Fixed RF trans-
mission IEC
61000-4-3
Conducted RF
IEC 61000-4-6
56
10 V/m
80 Mhz
2.7 GHz
3 V
0.15 MHz to
80 MHz 6 V in
ISM and am-
ateur radio
bandwidths
between
0.15 MHz and
80 Mhz
Confor-
The field intensities of
mant
fixed RF transmission
equipment outside
protected locations,
as determined in
an electromagnetic
Not
location test, should
applicable
not exceed 3 V/m.
Interference may
occur in proximity to
equipment labeled
as such:
distance. If the trans-
mission position is
known, the specific
distance can be calcu-
lated by means of
available equations.
See below for
calculation of
the separation