The device is set
to an intensity
level, but only mild
stimulation is felt
through the adhe-
sive electrodes.
The stimulation
current becomes
stronger, although
the device is set to
a low intensity.
The device stops
while in use.
There are changes
to or reddening of
the skin surface.
The batteries are
weak.
The surface of the
skin is not clean.
The entire adhe-
sive surface of the
electrode has no
adhesive effect and is
worn out.
The adhesive elec-
trodes are not fully
attached to the sur-
face of the skin.
The adhesive elec-
trodes only partially
adhere to the skin.
The batteries are
drained.
There is a fault in the
electronics.
This is possibly cause
by the adhesive
electrodes.
Technical malfunctions
Replace the batteries
with fully charged
batteries. Make sure
the polarity is correct.
Clean the skin.
The adhesive elec-
trodes must be
replaced.
Press the adhesive
electrodes firmly to
the skin's surface.
The adhesive elec-
trodes are old and
must be replaced.
Replace the batteries
with fully charged
batteries. Make sure
the polarity is correct.
Remove the batteries
and put them back in
after approx.
3 seconds.
Stop use immediately
and consult your
physician.