Table 2
Guidance and manufacturer's declaration – electromagnetic immunity
BC 60 is intended for use in the electromagnetic environment specified below. The customer or the user of BC 60 should assure that it is
used in such an environment.
Immunity test
Electrostatic discharge (ESD)
IEC 61000-4-2
Power frequency (50/ 60 Hz)
magnetic field IEC 61000-4-8
NOTE: U
is the a.c. mains voltage prior to application of the test level.
T
IEC 60601 test level Compliance level Electromagnetic environment – guidance
± 6 kV contact
± 6 kV contact
± 8 kV air
± 8 kV air
3 A/m
3 A/m
Floors should be wood, concrete or ceramic tile. If floors are
covered with synthetic material, the relative humidity should
be at least 30 %.
Power frequency magnetic fields should be at levels charac-
teristic of a typical location in a typical commercial or hospital
environment.