Electromagnetic Compatibility; Measuring Principle - Accu-Chek Aviva Manual Del Usuario

Ocultar thumbs Ver también para Aviva:
Tabla de contenido
Idiomas disponibles

Idiomas disponibles

Lancing device depths

Electromagnetic Compatibility

This meter meets the electromagnetic immunity requirements as per EN ISO 15197 Annex A. The chosen basis
for electrostatic discharge immunity testing was basic standard IEC 61000-4-2. In addition, it meets the
electromagnetic emissions requirements as per EN 61326. Its electromagnetic emission is thus low.
Interference from other electrically driven equipment is not to be anticipated.
Performance Analysis
The performance data for the ACCU-CHEK Aviva system (ACCU-CHEK Aviva meter with ACCU-CHEK Aviva test
strips) were obtained using capillary blood from diabetic patients (method comparison, accuracy), venous blood
(repeatability) and control solution (reproducibility). The system is calibrated with venous blood containing
various levels of glucose. The reference values are obtained using the hexokinase method. For method
comparison, the results were compared with results obtained using the hexokinase method with
84
deproteinization (automatic analyzer). The hexokinase method is traceable to an NIST standard.
The ACCU-CHEK Aviva system meets the EN ISO 15197 requirements.

Measuring Principle

Blood glucose concentrations may be measured in whole blood or plasma. Although you always apply whole
blood to the test strip, the meter displays blood glucose results that relate to plasma. Please refer to the test
strip package insert for information on how the system works, on the test principle, and on reference methods.
Protection class
III
Meter type
The ACCU-CHEK Aviva meter is suitable for continuous
operation
0.80, 0.95, 1.10, 1.25, 1.40, 1.55, 1.70, 1.85, 2.00, 2.15,
2.30 mm

Capítulos

Tabla de contenido
loading

Tabla de contenido