F1 - Scale Menu - Mettler Toledo ICS226 Instrucciones De Manejo

Tabla de contenido
Idiomas disponibles
  • ES

Idiomas disponibles

12
Settings in the menu
4.2 F1 – Scale menu
Factory settings are printed in bold characters.
4.2.1 F1.1 – Approval
No
OIML
NTEP
4.2.2 F1.2.1 – Calibration unit
Factory setting
Possible settings
4.2.3 F1.2.2 – Full capacity
Possible settings
4.2.4 F1.3 – Calibration
F1.3.1 – Geo value
Adaptation of the scale to the geographical location.
Factory setting
Possible settings
Note
Perform calibration
Note
No approval
Approval for OIML
Approval for NTEP
kg
kg, lb
3 kg, 6 kg, 15 kg / 6 lb, 15 lb, 30 lb
21
0 ... 31
Re-calibrate after changing Geo value.
1. E SCL appears, unload scale.
2. Start calibration by pressing
The scale determines the zero point.
The calibration weight to be placed on the scale is shown
on the display.
3. Place the calibration weight on the scale and confirm by
pressing
.
donE appears, calibration is completed.
In step 2, if you want to change the calibration weight, press
repeatedly until the desired calibration weight appears on
the display.
Press and hold
to abort the calibration process.
.
ICS226 30232252A
Tabla de contenido
loading

Tabla de contenido