30232252A ICS226
F1.3.3 – Linearity calibration
Perform calibration
4.2.5 F1.4 – Zero
F1.4.1 – Automatic zero setting
Factory setting
Possible settings
F1.4.2 – Power up zero
2%*
20%
* For approved mode, the setting is always 2%.
F1.4.3 – Pushbutton zero
Factory setting
Possible settings
* For approved mode, the setting is always 2%.
4.2.6 F1.5 – Tare
F1.5.1 – Auto tare
Factory setting
Possible settings
F1.5.2 – Auto clear tare
Factory setting
Possible settings
F1.5.3 – Pushbutton tare
Factory setting
Possible settings
1. E SCL appears, unload scale.
2. Start calibration with
.
After capturing the zero point, the first weight is shown on
the display.
3. Place the first weight on the scale and press
The first weight is confirmed. The second weight is shown
on the display.
4. Remove the first weight and place the second weight on
the scale and press
.
donE appears, calibration is completed.
0.5d
Off, 1d, 2d, 5d, 10d
Power up zero within a range of +/-2%
Power up zero from -2% ~ %18
2%*
2%, 20%, Off
Off
On, Off
Off
On, 9d
On
Off
.
13
Settings in the menu