1 GHz Spectrum Analyzer HM5510
All data valid at 23 °C after 30 minute warm-up
Frequency Characteristics
Frequency Range:
0.15 MHz...1.05 GHz
Stability:
±5 ppm
Aging:
±1 ppm/year
Frequency Resolution:
1 kHz (6
Center Frequency Range:
0...1.05 GHz
LO Frequency Generation:
TCXO with DDS (Digital Frequency Synthesis)
Span Setting Range:
Zero-Span and 1 MHz...1000 MHz
(1-2-5 Sequence)
Marker:
Frequency Resolution:
1 kHz, 6
Amplitude Resolution:
0.5 dB, 3
Resolution Bandwidths
(RBW) @3 dB:
500 kHz and 20 kHz
Video filter (VBW):
4 kHz
Sweep Time:
20 ms
Amplitude Characteristics (Marker Related) 150 kHz...1 GHz
Measurement Range:
-100...+10 dBm
Scaling:
10 dB/div.
Display Range:
80 dB (10dB/div.)
Amplitude Frequency Response (at 10 dB Attn., Zero Span and RBW
500 kHz, Signal – 20 dBm):
± 3 dB
Display (CRT):
8 x 10 division
Amplitude Scale:
logarithmic
Display Units:
dBm
Parameter Display (LCD):
2 Lines x 20 Characters, Center Frequency,
Span, Marker Frequency, Reference Level,
Marker Level
Input Attenuator Range:
0...40 dB (10 dB increments)
Tolerance of input attenuator: ± 2 dB relative to 10 dB position
Max. Input Level (continuous)
10...40 dB attenuation:
+ 20 dBm (0.1 W)
0 dB attenuation:
+ 10 dBm
Max. DC Voltage:
± 25 V
Max. Reference Level:
-50...+ 10 dBm
Reference Level Accuracy rel. to 500 MHz, 10 dB Attn., Zero Span and
RBW 500 kHz:
± 2 dB
Min. Average Noise Level:
approx. -100 dBm (RBW 20 kHz)
Intermodulation Ratio
typical › 75 dBc (2 Signals: 200 MHz,
rd
(3
Order):
203 MHz, -3 dB below Reference Level)
Harmonic Distortion Ratio
nd
(2
harm.):
typical › 75 dBc (200 MHz, Reference Level)
Bandwidth Dependent Amplitude Error rel. to RBW 500 kHz and Zero
Span:
± 1 dB
Inputs / Outputs
Measurement Input:
N-socket
Input Impedance:
50 Ω
VSWR: (Attn. ≥ 10 dB)
typ. 1.5 : 1
Supply Voltage
for Probes (HZ530):
6 V DC
Audio output (phone):
3.5 mm Ø jack
Test Signal output:
N-socket, output Impedance 50 Ω
Frequency:
10 MHz
Level
0 dBm (± 3 dB)
Functions
Keyboard Input:
Center Frequency, Reference Level,
Rotary Encoder Input:
Center Frequency, Reference Level,
Testsignal output Level, Marker,
Intensity (CRT), Contrast (LCD)
w w w . h a m e g . c o m
HM5510E/190107/ce · Subject to alterations · © HAMEG Instruments GmbH · ® Registered Trademark · DQS-certified in accordance with DIN EN ISO 9001:2000, Reg.-No.: DE-071040 QM
HAMEG Instruments GmbH · Industriestr. 6 · D-63533 Mainhausen · Tel +49 (0) 6182 800 0 · Fax +49 (0) 6182 800 100 · www.hameg.com ·
[email protected]
1
⁄
-digit in readout)
2
1
⁄
-digit,
2
1
⁄
-digit
2
A Rohde & Schwarz Company
General information
CRT:
D14-363GY, 8 x 10 div. with internal graticule
Acceleration Voltage:
approx. 2 kV
Trace Rotation:
adjustable on front panel
Power Supply:
105...253 V, 50/60 Hz ± 10 %, CAT II
Power Consumption:
approx. 31 W at 230 V/50 Hz
Safety class:
Safety class I (EN61010-1)
Operating temperature:
+5...+40°C
Storage temperature:
-20...+70°C
Rel. humidity:
5...80% (non condensing)
Dimensions (W x H x D):
285 x 125 x 380 mm,
with adjustable, lockable tilt handle
Weight:
approx. 5.6 kg
Accessories supplied: Line Cord, Operators Manual, x HZ21 Adapter Plug (N-
plug with BNC socket)
Optional accessories:
HZ20
Adapter, BNC to 4mm banana
HZ33
Test cable 50Ω, BNC/BNC, 0,5m
HZ34
Test cable 50Ω, BNC/BNC, 1m
HZ43
19''-Rackmount Kit 3RU
HZ520
Antenna
HZ525
50Ω-Termination, N plug
HZ530
Near Field Probe Set for EMI Diagnosis
HZ560
Transient Limiter
HZ575
Converter 75Ω to 50Ω
HZO30
Active probe 1GHz (0,9pF, 1MΩ, including many accessories)
S p e c i f i c a t i o n s
Subject to change without notice
25