Maintenance Tasks; Calibrating The Flow Rate; Replacing The So 3 Filter; Storage - Dräger X-act 5000 Instrucciones De Uso

Ocultar thumbs Ver también para X-act 5000:
Tabla de contenido
Idiomas disponibles
  • ES

Idiomas disponibles

5.3

Maintenance tasks

5.3.1

Calibrating the flow rate

NOTICE
i
i
For flow calibration, Dräger calibration tube X-act 5000
(8103651) has to be used. Observe the supplied instructions
for use 9022933.
The calibration will be done using a flowmeter.
1. Select "CALIBRATE FLOW" in the password-protected menu.
2. Follow the instructions on the display.
3. If necessary, set the required values using the arrow keys and
acknowledge the settings using the "OK" key.
Dräger X-act 5000
5.3.2
Replacing the SO
filter
3
1. Remove the 2 screws and the SO
2. Replace the SO
filter (2) with a new one.
3
3. Put the SO
filter door back in place and fasten the screws.
3
4. If necessary, enter the date of the filter change in the password-
protected menu "CHANGE SO3 FILTER".
5. If "BARCODE OPERATION AIR" is selected, the Dräger
X-act 5000 automatically tracks the amount of SO
reset the SO
counter in the password-protected menu.
3
6

Storage

Store the Dräger X-act 5000 dry and free of dust. Keep from direct
sunlight or heat radiation.
Storage
filter door (1).
3
. In this case,
3
2
1
00821900.eps
23
Tabla de contenido
loading

Tabla de contenido