During Use; To Be Considered During Use - Dräger X-act 5000 Instrucciones De Uso

Ocultar thumbs Ver también para X-act 5000:
Tabla de contenido
Idiomas disponibles
  • ES

Idiomas disponibles

3.1.5
Leak test
Dräger strongly recommends performing a leak test prior to safety-
relevant measurements and recommends against use if a leak test
fails. There are two options for a leak test:
Every time when the instrument is switched on, the leak test will be
optionally offered.
The leak test can be mandatory if set in the password-protected
menu.
The setting of the delivered Dräger X-act 5000 is that the
mandatory leak test is to be performed after 1000 strokes
(=100 L).
If the leak test is mandatory, "LEAK TEST" is indicated in the upper row
of the matrix display and at the right side of the segment display the
icon ("Warning") is indicated. The Dräger X-act 5000 cannot be
used until the leak test is passed.
The user will be prompted to start the leak test every time the pump is
switched on. The leak test may be skipped unless it is required due to
the settings in the password-protected menu.
If the leak test is selected, the user will be prompted to insert a closed
tube and start the test. When the leak test is passed, the green LEDs
will flash. When the "OK" key is pressed, the modes of operation will
be offered.
Dräger X-act 5000
NOTICE
i
i
If the leak test failed, the measuring modes will be inactive.
The instrument will display "LEAK TEST ERROR" and the
icon ("Error"). The red LEDs will flash.
If the leak test failed, the user can
repeat the leak test,
replace the SO
filter (see chap. 5.3.2 on page 23),
3
repair the instrument.
3.2

During use

3.2.1

To be considered during use

The Dräger X-act 5000 guides the user via menus. To use the
instrument, follow the guidance notes on the display.
To switch back to a previous option, use the EXIT function of the
instrument as follows:
Press the "UP" and "DOWN" keys simultaneously for 1 second.
The EXIT function is not available in the following cases:
during a self test,
during a leak test,
during a measurement,
during a calibration in the password-protected menu,
during an update of the tube database in the password-protected
menu.
Use
15
Tabla de contenido
loading

Tabla de contenido