Print-1 And Print-2 - OHAUS Adventurer Pro AV53 Manual De Instrucciones

Ocultar thumbs Ver también para Adventurer Pro AV53:
Tabla de contenido
Idiomas disponibles

Idiomas disponibles

EN-22
3.7.5 Unit (Cont.)
Defining Custom Unit
Set Custom to ON in the Unit menu to enable
and define the Custom Unit. The custom unit
is defined by a conversion factor and a least
Significant Digit (LSD). The conversion factor
is used by the balance to convert grams to the
custom weighing unit and is defined by entering
a factor and an exponent. The factor is a value
between 0.1000000 and 1.999999 inclusive.
The exponent moves the decimal point of the
factor to the right for positive values or to the
left for negative values.
Enter the Factor, following the instructions for
Editing Limits in Section 3.5.4. Enter the
Exponent and LSD using the Yes No buttons.
Factor
Exponent
(+3 to -3)
.1234
3
.1234
2
.1234
1
.1234
0
.1234
-1
.1234
-2
.1234
-3
Custom Unit = Conversion Factor x Grams.
The LSD is the value by which the displayed
weight is incremented or decremented.
LSD
Result
.5
Adds one decimal place
Display counts by 5
1
Display counts by 1
2
Display counts by 2
5
Display counts by 5
10
Display counts by 10
100
Display counts by 100
End Unit
Press the Yes button to advance to the Print-1
submenu. Press No to return to the first
available unit.

3.7.6 Print-1 and Print-2

The Print-1 submenu is used to set printing
parameters for an external printer or computer.
Print-2 contains an identical menu and is used
when the optional COM2 port is installed.
Output
Auto PrintOff, Cont., Interval, When Stable
Content Num Only-Off, all others
Layout
List
Conversion
End Print-1
Factor
123.4
Output
12.34
Set When Stable to On to print only stable
1.234
values. Set When Stable to Off to print stable
.1234
or unstable values.
.01234
.001234
Set GLP Tare to On to print GLP data once after
.0001234
a tare operation. Set GLP Tare to Off to disable
this feature.
Auto Print
When set to Continuous, the displayed value
is printed continuously. When set to Interval,
the displayed value is printed at the user
specified time interval (1 to 3600 seconds).
If set to When Stable, the balance will
automatically print the displayed value when
stability is achieved. An additional setting must
be made to determine if only stable non-zero
values will be printed (load setting) or if stable
zero and non-zero values will be printed (Load
& Zero). When set to Off, the Auto print feature
is disabled.
Adventurer Pro
PRINT-1
When Stable
On/Off
GLP Tare
On/Off
1 -3600 seconds
Stable- Load, Load & Zero
On-Header,Gross,Net, Tare,
Ref, Result,GLP
Line Format -Multi, Single
4LF-Yes/No
Form Feed-Yes/No
Yes/No

Hide quick links:

Tabla de contenido
loading

Tabla de contenido