8.3 Cavitron 30K Ultrasonic Inserts
The many styles of
Dentsply
applications.
Insert Tip
Shape and size
of tip determine
access and
adaptation.
Preheated lavage is
directed to the tip.
Hold the handpiece in an upright position. Activate to fill the handpiece with water. Lubricate the rubber O-ring on
the insert with water before placing it into the handpiece. Fully seat the insert with a gentle push-twist motion. DO
NOT FORCE.
88016 Rev. 5 (0119).indd 9
Cavitron ultrasonic inserts are easily interchangeable for various procedures and
O-Ring
Provides seal for handpiece coolant.
Finger Grip
Insert Marking
Manufacturer, Date (YYDDD = two digit year
and three digit day of year), Frequency, and
Type (e.g.,
17346 30K FSI-SLI-10S-FG)
DS
Connecting Body
Transfers and amplifies mechanical
motion of the stack to insert tip.
Magnetostrictive Stack
Converts energy provided by
the handpiece into mechanical
oscillations used to activate the
insert tip.
ENGLISH • 9