Scope Of Delivery; Instructions For Use (Qualified Personnel); Preparation For Use; Required Tools - Ottobock 2R67 Instrucciones De Uso

Tabla de contenido
Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 44
CAUTION
Mechanical overload
Falling due to breakage of load-bearing components.
► Do not subject the system components to mechanical vibration or
impacts.
► Check the product for visible damage before each use.
CAUTION
Incorrect setting of the torsion moments in the torsion unit
Falling due to unexpected behaviour of the product.
► The marking on the Allen head screw may not be turned as far as the
red area or beyond the red area (see fig. 1).

5 Scope of delivery

1 pc. 647G23 instructions for use (qualified personnel)
1 pc. 4R39/4R40/4R85/4R86 torsion adapter / 2R67 tube adapter with
torsion unit

6 Preparation for use

CAUTION
Incorrect alignment or assembly
Falling due to damage to the prosthesis components.
► Observe the alignment and assembly instructions.
CAUTION
Improper assembly of the screw connections
Falling due to breakage or loosening of the screw connections.
► Clean the threads before every installation.
► Apply the specified installation torque values.
► Observe the instructions for securing the screw connections and the
use of the correct length.

6.1 Required Tools

719R2 tube cutter
718R1 tube deburrer
710D4 torque wrench (up to 30 Nm)
710D12 torque wrench (up to 12 Nm)
18 | Ottobock