Hach POLYMETRON Serie Manual Del Usuario página 8

Tabla de contenido
Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 47
Figure 5 8315 Probe
Installation example for the 8394 probe
This probe installs perfectly in a Tri-Clover
also at a 90° angle starting from 2 inches (B). All Tri-Clamp
standards for Cleaning In Place (CIP).
Figure 6 8394 Probe
5.3.2 In a bypass
POLYMETRON flow chambers are designed not to retain air bubbles. Minimum recommended flow
rate: 100 mL/minute (6 L/hour) with homogeneous sample flow or ideally 330 mL/minute (20 L/hour).
Note: The progressive accumulation of bubbles on the surface of the probe reduces the active surface, increases
the cell constant, and leads to an abnormally low conductivity measurement.
8 English
®
Tri-Clamp
Tee starting from 1.5 inch diameter (A), and
Tees are in compliance with 3A
Tabla de contenido
loading

Tabla de contenido