Electromagnetic Immunity - Medela Swing Maxi Manual De Instrucciones

Ocultar thumbs Ver también para Swing Maxi:
Tabla de contenido
Idiomas disponibles

Idiomas disponibles

Electromagnetic Compatibility (EMC, IEC 60601-1-2:2007, Table 2)

Electromagnetic immunity

The electric breastpump Swing Maxi is intended for use in the electromagnetic environment
specified below. The customer or the user of the electric breastpump Swing Maxi should
assure that it is used in such an environment.
Immunity test
Electrostatic
discharge ( ESD )
IEC 61000-4-2
Electrical fast
transient / burst
IEC 61000-4-4
Surge
IEC 61000-4-5
IEC 60601
test level
± 6 kV contact
± 8 kV air
± 2 kV for power
supply lines
± 1 kV for input /
output lines
± 1 kV line( s ) to
line(s)
± 2 kV line( s ) to
earth
Electromagnetic
Compliance
environment
level
– guidance
Floors should be wood,
concrete or ceramic
± 6 kV contact
tile. If floors are covered
± 8 kV air
with synthetic material,
the relative humidity
should be at least 30%.
± 2 kV for power
Mains power quality
supply lines
should be that of a
± 1 kV for input/
typical commercial or
output lines
hospital environment.
± 1 kV differential
Mains power quality
mode
should be that of a
± 2 kV common
typical commercial or
mode
hospital environment.
89
Tabla de contenido
loading

Este manual también es adecuado para:

Swing maxi

Tabla de contenido