EMC / Technical Description
Electromagnetic Compatibility (EMC, IEC 60601-1-2:2001)
IEC 60601-1-2, Table 201
Electromagnetic Emissions
The Pump In Style
®
customer or the user of the PNSA should assure that it is used in such environment.
Emission Tests
RF emissions
CISPR 11
RF emissions
CISPR 11
Harmonic emissions
IEC 61000-3-2
Voltage fluctuations /
flicker emissions
IEC 60000-3-3
PNSA
Warning: The
should not be used adjacent to or stacked with other equipment. If adjacent or stacked use is necessary, the
should be observed to verify normal operation in the configuration in which it will be used.
The PNSA was not tested for immunity to electromagnetic disturbances.
56
All manuals and user guides at all-guides.com
Advanced (PNSA) is intended for use in the electromagnetic environment specified below. The
Compliance
Group 1
Class B
Class A
Complies
Electromagnetic environment - guidance
The PNSA uses RF energy only for its internal function. Therefore,
its RF emissions are very low and are not likely
to cause any interference in nearby electronic equipment.
The PNSA is suitable for use in all establishments, including
domestic establishments and those directly connected
to the public low-voltage power supply network that supplies
buildings used for domestic purposes.
PNSA