Settings
Code C8
Selecting the limit (threshold) value for suppression
of minor unbalance readings in grammes, or
ounces. The unit (gms or oz) depends on the
settings made under C3.
Ounces:
Grammes:
Range 0.12 to 0.71 oz
Range 3.50 to 20.0 g
Factory-adjusted to 0.18* oz
Factory-adjusted to 3.5 g
Select another limit, e. g.: 0.50 oz
Read out limit, e. g. 3.50 g
• Select Code C8
Select another limit, e. g. 5.50 g
• Set the value 0.50
• Press ENTER
Ounces:
Grams:
Range 0.25 to 2.00 oz
Range 3.50 to 20.0 g
Factory-adjusted to 0.25 oz
Factory-adjusted to 5.0* g
Read out limit, e. g. 0.25 oz
Select another limit, e. g.: 5.50 g
Select another limit, e. g. 0.50 oz
• Select Code C8
• Set the value 5.50
• Press ENTER
The selected mode of operation can be transferred to
the permanent
memory.
Code C10
Storing the selected mode of operation in the
permanent memory
0* =
No storage
1 =
Data are stored in the permanent memory
Store in the permanent memory – a three-tone
signal is given to acknowledge acceptance.
If the mode stored in the permanent memory is to be
changed, enter the desired state (e. g. on or off) for
the mode in question and transfer it via code C10 into
the permanent memory (not possible for code C4).
Code C12
Readings of numbers of measuring runs
Example: 222,123 measuring runs so far performed
(Fig. 7-3).
The following counters can be displayed:
1 = Total number of measuring runs
2 = Number of measuring runs where balance
quality was considered OK
3 = Number of optimisations or minimisations