Problem
No liquid out ows
from the insert during
operation
The device works
properly but the pump
strains/stresses
The pump turns
correctly, but when it
stops,
liquid out ows from
the handpiece
Insuf cient
performance
Possible cause
The insert type does not
foresee liquid passage
The insert is obstructed
The handpiece is obstructed
The irrigation level on the
screen is adjusted on "0"
The liquid bag is empty
adjuster has not been opened
The silicon tubing are
incorrectly installed
Excessive pressure of the
impeller on the peristaltic
pump tube
The lid of the peristaltic pump
is not closed correctly
The insert is not correctly
tightened on the handpiece
Insert broken, worn-out, or
deformed
09
TROUBLESHOOTING
Solution
Use an insert type with liquid
passage
Unscrew the insert from the
handpiece and free the insert
water passage by blowing
compressed through it.
If the problem persists, replace
the insert with a new one
Contact an Authorized Mectron
Service Center
Adjust the irrigation level
Replace the bag with a full one
adjuster
Check the connections
of silicon tubing
Check that the tube
inside the peristaltic pump is
properly positioned
(See paragraph 04.3)
Check that the lid of the
peristaltic pump is perfectly
closed (See paragraph 04.3)
Unscrew and correctly screw the
insert again with the Mectron
torque wrench
Replace the insert with a new
one