Mectron Piezosurgery white Manual De Uso página 24

Ocultar thumbs Ver también para Piezosurgery white:
Tabla de contenido
Idiomas disponibles
  • ES

Idiomas disponibles

NOT CORRECT
After each use of the PIEZOSURGERY
circuit (PUMP/FLUSH function) using an enzymatic detergent solution, to be followed by a
rinse with distilled water (see the Cleaning and Sterilization Manual).
Switch the device off by means of the O/I switch.
Disconnect all accessories before proceeding with their cleaning and disinfection.
CORRECT
white run a complete cleaning cycle on the irrigation
®
Put the insert inside the wrench as shown;
5
Firmly hold the central body of the
6
handpiece;
CAUTION:The handpiece must not
be grabbed by its terminal part and/or
cord, but only by its central body.
The handpiece must not be rotated, but
only rotate the wrench.
Rotate the wrench clockwise until the
friction snaps (the external body of the
wrench rotates compared to the handpiece,
The insert is now perfectly tightened;
On the keyboard, select the type of
7
function and irrigation necessary.
CAUTION: For the correct setting
of the Function parameters with the
the Chart annexed to this manual titled
"Appropriate settings for the inserts
on the PIEZOSURGERY
you've purchased.
05
USE
white" or the
®
21
EN