Quick Solution To Problems - Mectron Piezosurgery white Manual De Uso

Ocultar thumbs Ver también para Piezosurgery white:
Tabla de contenido
Idiomas disponibles
  • ES

Idiomas disponibles

09.2

QUICK SOLUTION TO PROBLEMS

Problem
EN
The device does not
turn on after having
brought the switch into
position "I"
The device is on
but not working. On
the display no error is
signalled
The device is on but
not working. One of the
following symbols is
active on the screen:
A slight whistling
sound coming from
the PIEZOSURGERY
handpiece is heard
during operation.
30
Possible cause
The electrical power cable
terminal is badly inserted in
the rear plug of the device
The electrical power cable is
defective
The fuses are out of order
The foot pedal plug is
incorrectly inserted in the
device socket
The foot pedal does not work
See paragraph 09.1 for the
possible cause, according
to the symbol active on the
display
The insert is not correctly
tightened on the handpiece
®
The irrigation circuit has not
Solution
Check that the power supply
Check that the power supply
socket works properly.
Replace the electrical power
cable
Replace the fuses
(see paragraph 09.3)
Correctly insert the pedal plug
in the socket on the back of the
device
Contact an Authorized Mectron
Service Center
See paragraph 09.1 for the
action to undertake, according
to the symbol active on the
display
Unscrew and correctly screw the
insert again with the Mectron
torque wrench
(see paragraph 05.4)
Fill the irrigation circuit via the
PUMP function
(see paragraph 05.4)
Tabla de contenido
loading

Tabla de contenido