Automatic interval mode can be interrupted via standby and analyses
started manually.
Operating
mode
Standby
The instrument is switched on and in standby.
Short press key 3 (pause) to switch standby on/off.
Note: During an interval pause, the instrument
immediately switches to standby; a started analy-
sis is always completed.
Manual
Precondition: The instrument is in standby or in an
interval pause.
Short press key 1 (manual) to switch on manual
mode.
An analysis is started immediately, irrespective of
the set interval pause.
Analysis process
The analysis takes approx. 2 minutes. The water analysis process for
determining the residual total hardness is as follows:
Analysis start
The solenoid valve for water inlet opens
The measuring chamber is flushed
The solenoid valve closes after the flush time
The optical transmitter is switched on to measure possible
turbidity in the water
The measurement continues until the turbidity clears (the wa-
ter has been fully outgassed)
The measurement starts when the measured value is con-
stant
The dosing pump and stirring bar are activated
The current measured value is read
The colour change in the measuring chamber is evaluated
If necessary, switch relay, result is displayed
The measuring chamber is flushed
Analysis end