Heyl Testomat 808 Manual De Instrucciones página 27

Dispositivo automático de análisis online para la dureza del agua
Tabla de contenido
Idiomas disponibles

Idiomas disponibles

Switch
Relay 2
position
(message poor analysis)
Normally open – closes in
0
case of 1 poor analysis,
remains closed until the next
good analysis
Normally open – closes in
1
case of 1 poor analysis,
remains closed until the next
good analysis
Normally open – closes after
2
2 poor analyses, remains
closed until the next good
analysis
Normally open – closes after
3*
3 poor analyses, remains
closed until the next good
analysis
Normally open – closes in
4
case of 1 poor analysis and
remains closed.
Opens again once the alarm
has been deleted
Normally open – closes after
5
2 poor analyses and re-
mains closed.
Opens again once the alarm
has been deleted
Normally open – closes after
6
3 poor analyses and re-
mains closed.
Opens again once the alarm
has been deleted
Normally open – closes in
7
case of 1 poor analysis,
remains closed until the next
good analysis
Normally open – closes after
8
the 2 poor analyses for 1
minute
Normally open – closes after
9
the 3 poor analyses for 1
minute
The following switch positions are possible:
Relay 1
(external flush)
Normally open – closes after a
poor analysis for 1 minute
Normally open – closes after a
poor analysis for 3 minutes
Normally open – closes after
the first poor analysis for 1
minute and after the second
poor analysis for 2 minutes
Normally open – closes after
the first poor analysis for 1
minute, after the second poor
analysis for 2 minutes and
after the third poor analysis for
3 minutes
Normally open – closes after
the first poor analysis for 1
minute
Normally open – closes after
the first poor analysis for 1
minute and after the second
poor analysis for 2 minutes
Normally open – closes for
90 s before the analysis
Normally open – closes after
the first poor analysis for 1
minute and after the second
poor analysis for 2 minutes
Normally open – closes after
the first poor analysis for 1
minute, after the second poor
analysis for 2 minutes and
after the third poor analysis for
3 minutes
* Delivery status
Description of the relay outputs
Comment
for relay 2
Continuous contact can be deleted
via key 4 (horn) or "Delete external-
ly".
Continuous contact can be deleted
via key 4 (horn) or "Delete external-
ly".
Continuous contact can be deleted
via key 4 (horn) or "Delete external-
ly". The instrument skips the interval
pause after the first poor analysis.
Continuous contact can be deleted
via key 4 (horn) or "Delete external-
ly". The instrument skips the interval
pause after the first and second poor
analysis.
Instrument goes into standby. Con-
tinuous contact can be deleted via
key 4 (horn) or "Delete externally".
Instrument goes into standby. Con-
tinuous contact can be deleted via
key 4 (horn) or "Delete externally".
The instrument skips the interval
pause after the first poor analysis.
Instrument goes into standby. Con-
tinuous contact can be deleted via
key 4 (horn) or "Delete externally".
The instrument skips the interval
pause after the first and second poor
analysis.
Continuous contact can be deleted
via key 4 (horn) or "Delete externally"
The instrument skips the interval
pause after the first poor analysis.
The instrument skips the interval
pause after the first and second poor
analysis..
Updated: July 2014