Error/Display
Error: Flap not closed
after blood application
Error: Internal error
(example 142)
Description/Solution
The measurement chamber flap was not closed properly after blood
application; the test did not start in time.
Solution
Remove the strip and repeat the test using a new test strip.
■
The meter has detected an internal error.
Solution
Power the meter off and on again. If the error still persists, the
■
meter is defective. Please call your local customer support and
service centre (see page 81).
Troubleshooting
77