3 Operation

3
Operation

Program Instrument and Turbine Maintenance

Switch on power switch.
LED
(Stand by) illuminates green.
Open front door.
Plug instruments or turbines on the provided couplings and
push back the flap (instruments must engage).
Turbines and contra-angles have to be inserted with the
chuck opening facing to the front.
Close front door.
Press button
(Start) on operating foil keypad
Program runs automatically. LEDs 1,2,3 of the loaded places
toggle during cycle.
All attached instruments or turbines are lubricated in
succession.
Following maintenance of attached instruments or turbines, all
LEDs of loaded places will illuminate.
Open front door.
Remove instruments or turbine (for instruments loosen
engaging lever).
LEDs will turn off upon removal of instruments.
Check attachment handpiece adapter to adapter holder
The attachment of the adapter to the adapter holder
be verified at least once per week.
If needed, the coupling nut should be tightened

Program Chuck Maintenance

Switch on power switch.
LED
(Stand by) illuminates green.
Open front door.
Press instrument with chuck opening to maintenance jet.
Press
button on operating foil keypad.
Program runs automatically.
LED
will illuminate during the process.
Remove instrument from jet when maintenance is complete
and push back the flap