Plug connection with integrated low-voltage supply (15 V ⎓) for the TM 50 Stirrer incorporated
Stirrer/pump:
in the bottom of the casing of the TITRONIC
Housing:
Material:
Polypropylene
Front keyboard: polyester coated
Dimensions:
13.5 x 31 x 20.5 cm (W x H x D), height incl. interchangeable unit
Weight:
approx. 2.0 kg for basic unit
Ambient conditions:
Climate:
Ambient temperature: + 10 ... + 40 °C for operation and storage
Humidity according to EN 61 010, Part 1:
Max. relative humidity 80 % for temperatures up to 31 °C,
linear decrease down to 50 % relative humidity at a temperature of 40 °C
Altitude:
Device: No restrictions
Power supply: up to 5000 m
Pollution degree:
Pollution degree IP 20, indoor use only.
Dosing units:
Cylinder:
20 ml and 50 ml, borosilicate glass 3.3 (DURAN
Valve:
volume neutral cone valve made from fluorocarbon polymers (PTFE), TZ 3000
Hoses:
FEP hose set, blue
Dosing accuracy:
after DIN EN ISO 8655, part 3:
Accuracy:
Precision:
Do not use the device in hazardous locations!
0.15 %
0.05 %
®
300 piston burette
®
)
63