SPECIFICATIONS
Electrical
Temperature scale:
Celsius (℃), Fahrenheit (℉) or Kelvin (K) user-selectable
Measurement range:
-50℃ to 1300℃, -58℉ to 2000℉, 223K to 2000K
Resolution:
1℃ or 1℉, 0.1℃ or 0.1℉,1 K
Accuracy:
Accuracy is specified for operating temperatures over the range of 18℃
to 28℃(64℉ to 82℉), for l year, not including thermocouple error.
±2℃---------------------- -50℃ to 0℃
±4℉---------------------- -58℉ to 32℉
±5K---------------------- 223K to 273K
±(0.5% rdg+1℃)-------- 0℃ to 1000℃
±(0.8% rdg+1℃)-------- 1000℃ to 1300℃
±(0.5% rdg+2℉)-------- 32℉to 2000℉
±(1.0% rdg+2K)-------- 273Kto 2000K
Temperature coefficient:
0.1 times the applicable accuracy specification per ℃ from 0℃ to 18℃
and 28℃ to 50℃ (32℉ to 64℉ and 82℉ to 122℉).
Input protection:
60V dc or 24V rms ac maximum input voltage on any combination of
input pins.
Reading rate:
2.5 times per second.
Input connector:
Accepts standard miniature thermocouple connectors (flat blades
spaced 7.9mm, center to center).
2
Environmental
Ambient operating range:
0℃ to 50℃ (32℉ to 122℉)
Storage temperature:
-20℃ to 60℃ (-4℉ to 140℉)
Relative humidity:
0% to 80% (0℃ to 35℃) (32℉ to 95℉)
0% to 70% (35℃ to 50℃) (95℉ to 122℉)
General
Display:
3 /
1
digit liquid crystal display (LCD) with maximum reading of 1999
2
Battery:
Standard 9V battery (NEDA 1604, IEC 6F22)
Dimensions:
162mm (H)×76mm(W)×38.5mm(D)
Weight: 210g
Supplied probe:
1 x 1.2m type "K" thermocouple bead probe (Teflon tape insulated).
Maximum insulation temperature 260℃(500℉)
Probe accuracy±2.2℃ or ±0.75% of reading (Whichever is greater)
from 0°to 800℃