SETTING UP THE MYCARE INSITE ANALYSER AND TOUCH SCREEN
AVOID
• Condensing humidity and water
• Heat and large temperature fluctuations
• Electromagnetic radiation
Place the Analyser on a dry, clean, stable, and level surface.
Make sure the Analyser has at least 10 cm of table surface and clearance on each side, and that the Analyser can
be easily disconnected from the power source. Allow the Analyser to acclimate to the ambient room temperature
before operation.
Plug the Power Adapter (AC adapter plug) into the Analyser and the Power Cable into the wall socket.
The MyCare Insite is operated with a Touch Screen.
To ensure that the Analyser is properly grounded, use a 3-pole power outlet that includes neutral, hot, and ground
sockets. The MyCare Insite must be connected to a power outlet with a 3-pole plug.
A certified electrician can confirm that the grounding meets requirements.
Ensure the Analyser is plugged in. Connect the Touch Screen to the USB port of the Analyser and then turn on the
Touch Screen.
The MyCare Insite Launch Screen
1. Connection Screen
The connection screen appears when the Touch
Screen is connected to the Analyser.
Tap Next to continue.
PI MCI-EUR-ML-PKG Rev 01
English 6 / 12
• Electrostatic discharge
• Vibrations (e.g. from centrifuges and dishwashers)
• Direct sunlight
The launch screen appears by pressing
the Touch Screen home button or when
the Touch Screen is not connected to
the Analyser.
Touch the MyCare Insite logo to start
the MyCare Insite application.
Touch Browser to open the Android web
browser.
Touch Settings to open the Android
settings menu.
2. Main Menu
The MyCare Insite application starts and the main
menu appears.
Note: while the MyCare Insite is warming-up (indicated
by the flashing green LED on the front, see page 5) the
"Measurement" button is grey and inactive.