Service & maintenance information
Service and maintenance of a medical device is wholly the responsibility of the owner of that device.
Failure to service and maintain a device in accordance with the instructions may invalidate the
device's warranty. Furthermore, failure to service and maintain a device may compromise the clinical
condition or safety of users and/or their caregivers. Do not perform service and maintenance while the
user is in the product. Contact the local dealer for assistance, if needed, in setting up, using or how to
perform service and maintenance of the device.
Lifetime
The lifetime of this product in normal use is 5
years if all maintenance and servicing is carried
out in accordance with the manufacturer's
instructions and demonstrably recorded. The
device may be reissued within its lifetime.
Service interval
The
frequency
of
inspection
altered depending upon use and wear. It is
recommended that the product is inspected
yearly, every time the product is re-issued for
use and after a long-term storage (more than 4
months). The inspection should be carried out
by a person who understands the use of the
product.
Service checklist:
Check and adjust the following as per their
intended use:
• Backrest and leg support is adjustable
and can be locked in position
• Zipper(s) & Velcro
• Chair is foldable/unfoldable
• Nuts and bolts on the product are
tightened
• Product for cracks or signs for worn-out
parts
• All labels on the product are intact
Repair or replace any damaged or worn out
parts.
Maintenance
Before every use it is recommended to:
Wipe off any residue and dirt from the product,
using a cloth with warm water and a mild
detergent/soap without chlorine and let it dry
before use carry out a visual inspection, for any
damaged or worn-out parts.
Washing
Main product
This product can be washed at 60° using a
mild detergent for 10 minutes in a washing
machine designed for medical devices. Use the
machine's drying function to dry the product.
Unmount all accessories and wash them
separately.
This product can be washed at 85° using a
mild detergent for 3 minutes in a washing
machine designed for medical devices. Use the
machine's drying function to dry the product.
Unmount all accessories and wash them
separately.
EN
Fabric and covers
The cover can be washed at max. 60°C using a
mild detergent in a washing machine.
Handwash
This product and its accessories can also be
washed by hand. Use warm water and a mild
detergent/ soap without chlorine and let all
parts dry before use.
may
be
Disinfection
The product can be disinfected with a 70%
disinfectant IPA solution. It is recommended to
wipe off any residue and dirt from the product,
using a cloth with warm water and a mild
detergent/soap without chlorine and let it dry
before disinfection.
Materials
• Aluminium
• Rubber
• Plastic
• Polyester
• Stainless steel
Surface treatment
The following surface treatment processes
were used to protect against corrosion:
• Lacquered surfaces = Polyester powder
coating or ED-coating
• Non-lacquered aluminium parts =
Anodized coating
• Non-lacquered steel surfaces = Galvanized
21