What to do if ...
Measured value
unstable
Slope too low
Measuring range
Reproducibility
pH range
Temperature range
Membrane
resistance
Length
Diameter
22
What to do if ...
Cause
– Inner junctions not
sufficiently wetted with
bridge electrolyte (Ca 800)
– Junctions encrusted
– Cable broken
Cause
– Conditioning time too short
– Standard solutions too old
– Junctions encrusted
– Electrode defective
Technical data
0.02 ... 40,000 mg/l Ca
± 4 %
2.5 ... 11
(see I
NTERFERENCES
0 ... 40 °C
1 to 4 MΩ
170 mm (including 50 mm connection head)
Shaft:
12 mm
Connection head
16 mm
Remedy
– Fill up bridge electrolyte until the inner
junctions are covered with bridge
electrolyte
– Leave the bridge electrolyte to react on
the junctions for some hours until the
crusts have dissolved.
– Exchange the electrode
Remedy
– Extend the conditioning time
– Use new standard solutions
– Leave the bridge electrolyte to react on
the junctions for some hours until the
crusts have dissolved.
– Exchange the electrode
2+
-7
(5 x 10
... 1 mol/L Ca
)
Ca 800
2+
)
ba15352d04
12/2021