11. TROUBLESHOOTING
Check the following if your unit should fail during operation. You can also refer
to the pages of this manual for complete instructions.
Problem
Nothing happens
when the power
switch is pressed.
No nebulization or
low nebulization
rate when the
power is on.
The device is
abnormally loud.
The device is very
hot.
NOTE: If the suggested remedy does not solve the problem, do not try to
repair the device. No parts of the unit are user serviceable.
Return the unit to an authorized retail outlet or distributor.
Cause
The AC adapter is not
connected correctly to the
electrical outlet or the
compressor.
There is no medication in
the medication tank.
There is too much/little
medication in the
medication tank.
The vaporizer head is
missing or not assembled
correctly.
The nebulizer kit is not
assembled correctly.
The nozzle is blocked.
The nebulizer kit is tilted at
a sharp angle.
The air tube is not
connected correctly.
The air tube is folded or
damaged.
The air tube is blocked.
The air filter is dirty.
The air filter cover is not
attached correctly.
The compressor is covered
by something.
The device was operated
continuously for over 20
minutes.
English 9
Remedy
Check that the plug is inserted in
an electrical outlet. Unplug then
reinsert the plug if necessary.
Add the correct amount of
medication to the medication
tank.
Attach the vaporizer head
correctly.
Assemble the nebulizer kit
correctly.
Make sure that the nozzle is free
of blockages.
Make sure that the nebulizer kit
is not tilted at an angle of more
than 45 degrees.
Make sure that the air tube is
correctly connected to the
compressor and nebulizer kit.
Make sure that the air tube does
not contain kinks.
Make sure that the air tube is
free of blockages.
Replace the air filter with a new
one.
Attach the air filter cover
correctly.
Do not cover the compressor
with any type of cover during
use.
Limit use to 20 minutes at a
time, and allow a 40 minute rest
interval before using the device
again.