9. CLEANING
1. Turn the power off and unplug from the wall outlet.
2. Remove the tube from the air inlet.
3. TO CLEAN: Disassemble the mouthpiece, nebulizer cup, chamber and
baffle, and wash these items in hot water with dish-washing detergent.
Rinse these items thoroughly to remove the detergent and air dry.
4. TO DISINFECT: Mix one part of white vinegar with 3 parts hot water in a
clean container. Submerge mouthpiece, nebulizer cup, chamber and baffle
for half an hour in the solution. Remove from solution and let air dry.
5. There is no need to clean the tube. If necessary wipe the surface regularly.
Note:
Replace the nebulizer accessories including adult mask, child mask,
mouthpiece, nebulizer cup, Air tube after using 6 months.
Replace the filter every 30 days or if the filter turns grey.
CAUTION: The jet nebulizer must be changed if it becomes clogged.
CAUTION: The nebulizer and accessories must not be boiled.
10. MAINTENANCE
1. The air compressor filter should be replaced when it turns gray.
Additional filters can be procured from your dealer.
2. Only authorized personnel shall make repairs to this product.
CAUTION: The nebulizer may be damaged if used with a dirty filter or
if the filter is replaced by some other material such as
cotton. Do NOT operate without a filter.
English 8