Set Up Menu
Setup Menu
LCD Invert
Beep
Track
Zero
Calibrate
X
Setting Options
LCD Invert:
Invert (!) or non-invert (")
Beep:
Enable (!) or disable (")
Track:
Put the instrument into Track mode
Zero:
Reset the stored Zero value
Calibrate:
Put the instrument into Calibrate mode
X:
Exit without saving changes
Save changes and exit
Track Mode
Pressing and holding the ZERO button on 'Track' will put the instrument into Track mode. The instrument will
remain in Track mode until it is disabled in the set up menu, or is power cycled.
The instrument will always power up in its default mode (Click mode).
Zero Resetting
Zero resetting should only be performed if the checker won't zero. If this is the case the transducer may have
been overstrained.
If the reading is unstable the instrument should be returned to a supplier approved agent for repair. It is
recommended to get the instrument recalibrated as soon as possible after resetting the zero.
Make sure no torque is being applied to the checker and remove any tools from the input drive when
resetting the zero.
Calibrate Mode
Only to be used by a supplier approved agent.
USB Port
The USB port provides the power for the Digital Torque Checker.
The USB port can also be used to update the Digital Torque Checker software.
Press the ZERO button to select the menu item
!
!
"
Press and hold the ZERO button to confirm/toggle