Error Codes - IKA RV 10 basic Instrucciones De Manejo

Ocultar thumbs Ver también para RV 10 basic:
Tabla de contenido
Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 51
Any malfunctions during operation will be identified by an error message on the display (only RV 10 digital).
Once a serious error message has been displayed, the lift moves to the top end position and the device can no longer be operated. The lift can be
operated again.
Proceed as follows in such cases:
 GSwitch off device using the main switch
 Carry out corrective measures
 Restart device
Error
Effect
code
E01
No or too little rotation
E02
No communication with the
heating bath
If the actions described fail to resolve the fault or another error code is displayed then take one of the following steps:
- contact the IKA
service department,
®
- send the device for repair, including a short description of the fault.
RV 10.2029
RV 10.2030
RV 10.2031
RV 10.610
RV 10.606
RV 10.607
RV 10.608
RV 10.609
RV 10.2032
RV 10.2033
RV 10.2034
RV 10.2035
RV 10.3000
RV 10.8001
PC 1.1

Error codes

Cause
Cable break
Load too large
Motor blocked
Interface hidden or dirty
NS 24/40 Distillation spider with 6 sleeves
NS 24/40 Distillation spider with 12 sleeves
NS 24/40 Distillation spider with 20 sleeves
Distilling sleeve 20 ml
NS 29/32 Distillation spider with 5 flasks 50 ml
NS 29/32 Distillation spider with 5 flasks 100 ml
NS 24/29 Distillation spider with 5 flasks 50 ml
NS 24/29 Distillation spider with 5 flasks 100 ml
NS 29/42 Distillation spider with 5 flasks 50 ml
NS 29/42 Distillation spider with 5 flasks 100 ml
NS 24/40 Distillation spider with 5 flasks 50 ml
NS 24/40 Distillation spider with 5 flasks 100 ml
Extension plate
Seal, PTFE compound
Cable
Labworldsoft
®
Correction
Reduce the volume in the evaporator piston
Check / clean the interface
34
Tabla de contenido
loading

Este manual también es adecuado para:

Rv 10 digitalRv 10 basic vRv 10 basic vcRv 10 basic dRv 10 basic dcRv 10 digital v ... Mostrar todo

Tabla de contenido