Prestige MEDICAL OMEGA 121/13 Instrucciones De Funcionamiento página 8

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 46
GB
In the unlikely event that something should go wrong, we have incorporated a number of safety
features to ensure that your autoclave remains safe at all times.
Safety Features.
1. If for any reason, the temperature falls below
the minimum required sterilizing
temperature, resulting in the Sterilizing
Temperature light (L4) switching off, the cycle
timer will re-start from zero once the correct
temperature has been restored.
2. If there is an electrical or electronic failure
resulting in a build up of pressure - in excess
of normal operating pressure - one or all of
the following safety features will be activated:
i) The PRIMARY Pressure Release Valve will
loudly and rapidly "vent" steam.
ii) Should the PRIMARY Pressure Release
Valve fail to operate, a SECONDARY
release valve will operate at a slightly
higher pressure.
Technical Specifications
Height
420mm
Width
350mm
Net Weight
9.2 kilos
Temperature
Sterilizing time
Cycle time (nominal)
Minimum rest period
Fuses - Located on rear plate and in power unit (FS3), the
latter are NOT user replaceable item. Only qualified
personnel should attempt to change it. Rear plate
fuses (230v models only) F7.5A 32 x 6.3mm,
ceramic sand filled, FS3 on 230v model power unit pcb,
T500mA, to IEC 127 (120v, T500mA) Mains plug top fuse
(User replaceable), F13A to BS1362 UK ONLY. All models
are continuously rated for loading.
Body - Deep drawn aluminium. 3103-0
Lid - Gravity cast aluminium. LM 25,TF
Gauges - Depending on the cycle selected, the minimum
gauge readings will be as follows:
Temperature 121°C cycle - 121°C
Pressure
121°C cycle - 1.05 bar 134°C cycle - 2.05 bar
Heater - Externally surface mounted mechanically fixed
electric element.
Temperature Cut Out - Bi-metallic type with manual re-set
isolating at 170°C
Max. Single Fault Temperature - 140.5°C
Model information
Capacity
Internal chamber
Dimensions (d/h)
121˚C
134˚C
16 mins
3 mins 40 secs
28 mins
20mins
5 mins
5 mins
134°C cycle - 134°C
iii) A resettable thermal cutout located on the
base of the autoclave will 'switch' at a
predetermined temperature, disconnecting
the autoclave from the mains supply.
(see Performance Test Certificate).
Should any of the devices listed above activate,
please observe the following steps:
a) Do not touch the autoclave
b) Switch off at the wall socket and remove
plug from wall socket.
c) Wait until the autoclave has cooled down
before attempting to remove the lid.
d) Do not attempt to re-start a sterilizing
cycle.
e) Arrange for an immediate service call by
contacting the company from which the
unit was purchased.
10 litres
Max' Instrument
length
210/270mm
Max' Load Weight
Volts / watts / frequency
120v models / 1200 watts / 50 - 60 Hz
230v models / 1500 watts / 50 - 60 Hz
Pressure - Primary valve: dump type, manufactured to
Section Vlll, ASME Code of Practice, calibrated
and sealed. Operates at 2.65 bar (38.4 psi)
minimum, maximum accumulation plus 10%.
- Secondary valve: proportional type -
uncalibrated.
Over Voltage Category - Group ll
Pollution Degree - Group 2
Environment Conditions
- indoor use
- altitude up to 2000m
- temperature 5°C to 40°C
- maximum relative humidity 80% for
temperatures
up to 31°C decreasing linearly to 50%
relative
- mains supply voltage fluctuations not to
exceed +10% of the nominal voltage
Input Connections - Mains inlet socket 'hot' format
conforming to IEC 302.
Safety Shut Down - See 'Temperature Cut Out', which
should only be reset by a qualified engineer.
Packaging - All packaging materials are recyclable.
8
290mm
4.0 kilos
humidity at 40°C.
loading

Productos relacionados para Prestige MEDICAL OMEGA 121/13