GB
To ensure your autoclave gives you the years of service for which it was designed, it is important to
remember a few "do's" and "dont's" with regards to the operation of the unit and to carry out the
simple care and maintenance procedures on a weekly basis.
Do ensure that....
1...you read these instructions and always
follow the operating sequence.
2...the instruments are designed to withstand
the sterilizing temperatures selected, are
thoroughly cleaned and rinsed before
sterilizing, and are not any longer than
the length, or exceed the load weight,
specified - see "Technical data" section.
3...the water level is maintained regularly with
clean distilled or de-ionised water only.
4...the unit is in a "draft free" environment and
is positioned not less than 250mm from
adjacent walls.
5...you only use red sealing gasket (229216)
and that it is changed at the end of it's life,
if visibly damaged, or when shrinkage has
occurred. See "Fault mode-4"
6...the lid is securely closed when the unit is
not in use, to avoid the risk of accidental
damage. Never leave in position (S2).
7...you record the temperature and pressure
readings for each cycle completed, to ensure
they remain constant between cycles. Check
readings against those quoted - see
"Technical data" section.
8...you quote your model details, serial number
and date of purchase when contacting
Prestige Medical or your supplier.
Do not....
1...lose this operating instruction manual.
2...add any chemicals whatsoever to the water.
3...attempt to sterilize volatile substances,
toxic materials or inappropriate loads.
4...place the unit on heat sensitive surfaces
ie polished wood or glass.
5...open the Depressurization Valve (A) during
the sterilization cycle.
6...leave the Depressurization Valve (A) in the
"open" position when placing the lid upside
down on a work surface.
7...immerse the unit or electrical cord in water
when cleaning.
8...use abrasive materials or lubricants when
cleaning.
Continued Operation
9...drop or abuse the unit.
10..use in areas of risk associated with
11.attempt to change any fuses until the unit
12.remove the access panel on the lid for
13.reach over the unit when removing cover,
14.cover lid or obstruct steam venting area
* These operations should only be
undertaken by a qualified person.
Care and Maintenance.
1. Remove from inside the lid and clean with
2. Rinse thoroughly, shake dry, do not wipe.
3. Replace in the lid by tucking evenly under
4. Replace when it begins to show signs of
6. If a new gasket leaks, or a persistent leak
both the lid and body of the unit with a
"Scotchbrite" scrubbing pad making sure
7. Clean both interior and exterior with warm
8. Monitor the first cycle of the day to check
9. Prestige Medical recommend that your unit
10. Lubricate underside of body lugs with
6
flammable materials or gases.
has been unplugged from the mains*.
reasons other than servicing*.
to do so may cause burns from rising heat
and steam.
when manually depressurizing.
Red Sealing Gasket.
warm, soapy water.
all lugs. It may appear slightly wrinkled until
used.
leakage.
Autoclave.
develops, gently clean the sealing surface of
you do not remove any metal. Rinse both
surfaces, but do not dry.
soapy water ensuring electrical parts are
kept dry.
the Air Bleed Device which is located inside
the lid, audibly "clicks" shut.
is calibrated at six monthly intervals
"vaseline" if the lid becomes stiff.
DO NOT LUBRICATE GASKET